DONGGUAN LONROY EQUIPMENT CO LTD

Manufacturer from China
Verified Supplier
1 Years
Home / Products / Paint Testing Equipment /

BGD 872 ISO 15711 Cathodic Disbondment Test Chamber

Contact Now
DONGGUAN LONROY EQUIPMENT CO LTD
Visit Website
City:dongguan
Province/State:guangdong
Country/Region:china
Contact Person:MsKaitlyn Wang
Contact Now

BGD 872 ISO 15711 Cathodic Disbondment Test Chamber

Ask Latest Price
Video Channel
Brand Name :Lonroy
Model Number :BGD 872
Certification :Includes Calibration Certification
Place of Origin :China
MOQ :1
Price :Negotaible
Payment Terms :L/C,D/A,D/P,T/T,Western Union
Supply Ability :200
Delivery Time :5-8 work days
Packaging Details :Wooden package
Equipment Dimensions :127cm × 100cm × 120cm
Test Chamber Dimensions :73cm × 65cm × 73cm (1m³)
Power Supply :AC 220V/50Hz
Temperature Resolution :0.1℃
Warranty :1 Year
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

BGD 872 ISO 15711 Cathodic Disbondment Test Chamber

1.Description

The Cathodic Disbondment Test Chamber is used to evaluate the resistance of coatings (typically applied to protect metal substrates in ships and marine structures) to cathodic disbondment when exposed to seawater. It accurately assesses the adhesion failure between coatings and metal surfaces caused by cathodic protection, which often originates from defects in the coating system, such as accidental coating loss, application flaws, or excessive permeability.

Artificial holidays (defects) are created on coated specimens, which are then immersed in the test solution (artificial seawater) within the chamber. Two specimens are connected to a cathodic protection circuit (Method A) or sacrificial anode (Method B), while two additional specimens serve as controls without cathodic protection or sacrificial anode connection. After testing, the specimens are examined for blistering and underfilm penetration around the artificial holidays. Blistering and disbondment grades are evaluated according to relevant standards, and the coating's performance under cathodic protection conditions is determined by comparing protected and unprotected specimens.


2.Technical Parameter


Test Chamber Dimensions (L×W×H)

73cm × 65cm × 73cm (1m³)

Solution Temperature Range

20℃ ~ 80℃

Temperature Resolution

0.1℃

Temperature Uniformity

≤1.5℃

Temperature Fluctuation

±0.5℃

Temperature Control Accuracy

±0.1℃

Heating Method

Electric heating with PID+SCR balance control

Aeration Device

PP tube microporous bubbling

Power Supply

AC 220V/50Hz, Max. Current: ~10A, Max. Power: 3kW

Equipment Dimensions (L×W×H)

127cm × 100cm × 120cm

Standard Electrodes

Anode Electrode (Graphite) and Reference Electrode




Inquiry Cart 0