DONGGUAN LONROY EQUIPMENT CO LTD

Manufacturer from China
Verified Supplier
1 Years
Home / Products / Paint Testing Equipment /

High Low Temperature Humidity Cabinet Programmable Constant Temperature Humidity Device

Contact Now
DONGGUAN LONROY EQUIPMENT CO LTD
Visit Website
City:dongguan
Province/State:guangdong
Country/Region:china
Contact Person:MsKaitlyn Wang
Contact Now

High Low Temperature Humidity Cabinet Programmable Constant Temperature Humidity Device

Ask Latest Price
Video Channel
Brand Name :Lonroy
Model Number :BGD 897
Certification :Includes Calibration Certification
Place of Origin :China
MOQ :1
Price :Negotaible
Payment Terms :L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability :200
Delivery Time :5-8 work days
Packaging Details :wooden package
Dimensions of Machine :1000×1650×1250mm
Working Room Size :500×750×600mm
Power Supply :220V/380V; 50HZ
Total Power :5.5KW
Warranty :1 Year
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

BGD 897 High-Low Temperature & Humidity Cabinet Programmable Constant Temperature & Humidity Device

1.Description

These new programmable cabinets with a precise system of temperature and humidity control, which provide various necessary environmental simulative conditions for industrial research and biotechnology tests. Widely applied in sterile tests and stability check-up of pharmaceuticals, textile and food processing as well as tests in material performance, packing and lifetime of industrial products.

Biuged launched the latest generation climate chambers which has the following updates compared to previous products or competitors prodcuts:

1.Use electronic humidity sensors to control and display the relative humidity of the working room, more convenient , more accurate.

This new design avoids many inconveniences and issues come from the previous method of using PT 100 platinum resistance for wet and dry bulb measurements. For example there is no need to regularly replace the wet bulb fabric, and some users may forget to replace it or hang it incorrectly, resulting in wrong humidity readings or the inability to control humidity during the test.

2.The refrigeration control system adopts an electronic expansion valve to control the cold end balance, which, under the same test conditions, can save at least 40% power compared to products using thermal balance methods. It also reduces water consumption by more than 60%.

3.Using a scroll compressor results in a larger cooling capacity and lower noise.

2.Technical Parameter

Ordering
Information

Temperature
Range

Humidity Range

Total Power/Max. Current

Working Room Size
(
W×H×D)

Overall Size
(
W×H×D)

BGD 897/100B

-20150C

2098%

4.6KW/16A

400×500×500mm

900×1400×1150mm

BGD 897/100C

-40150C

BGD 897/100D

-60150C

BGD 897/225B

-20150C

5.5KW/22A

500×750×600mm

1000×1650×1250mm

BGD 897/225C

-40150C

BGD 897/225D

-60150C

BGD 897/608B

-20150C

11KW/12A

(380V; 3 phase and 4 wires)

800×950×800mm

1300×1850×1400mm

Temperature Uniformity

± 2 ℃

Humidity Uniformity

≤ 2%~3% RH(≧ 75%RH) or ≤±5% RH(≤75% RH)

Temperature Stability

±0.5 ℃

Humidity Stability

±2% RH

Temperature Increasing Rate

3.5℃/min(no-load, average value during the whole test)

Temperature Decreasing Rate

≧1℃/min(no-load, average value during the whole test)

Power Supply

220V/380V; 50HZ

Total Power

5.5 KW

Inquiry Cart 0