BGD 266 Integrated FID Gas Chromatograph Analyzer Machine

Ask Latest Price
Video Channel
Brand Name :Lonroy
Model Number :BGD 266
Certification :Includes Calibration Certification
Place of Origin :China
MOQ :1
Price :Negotaible
Payment Terms :L/C,D/A,D/P,T/T,Western Union
Supply Ability :200
Delivery Time :5-8 work days
Packaging Details :Wooden package
Size :500mm × 430mm × 440mm
Net Weight :50kg
Temperature Control Accuracy :≤0.1ºC
Warranty :1 Year
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

BGD 266 Gas Chromatograph

1.Description

Samples in gaseous, liquid, or solid forms are introduced into the chromatographic (separation) column system via a vaporization chamber. Under the carrier gas , the mixture components flow through the column, where they are separated by interactions with the stationary phase. The separated components are then carried into a detection system, where non-electrical signals (e.g., chemical composition) are converted into electrical signals (current or voltage) proportional to their concentrations. These signals are recorded or processed by a data system for qualitative and quantitative analysis of the mixture components.

Features:
* Cost-effective with high performance-to-price ratio
* Modular integrated design for easy installation and maintenance
* Compatible with chromatography data workstations to enhance workflow efficiency
* Hydrogen Flame Ionization Detector (FID) - High sensitivity for organic compound analysis, ideal for solvent residue detection in printed materials and solvent composition analysis.
* Capillary column inlet design
* Spacious column oven (300×300×200mm) - Ensures precise complex analyses, compatible with capillary columns.
* Advanced noise suppression system - Eliminates fan noise during operation for a quiet working environment.
* All-glass injection system with options for rapid vaporization or on-column injection.
* Dual-channel gas system with precision pressure stabilization - Ensures instrument reproducibility.
* Microcomputer-controlled operation with user-friendly interface for simplified workflows.
* Advanced temperature control system - Microcomputer-managed temperature settings for the column oven.
* Overheat protection - Programmable temperature limits up to 400°C with dual safeguards (software and hardware).
* Chromatography workstation integration - Enables extensive data storage and provides specialized analytical methods.

2.Technical Parameter

Temperature Control Range RT +10°C - 400°C, 8-step programmed temperature ramp
Heating Rate 0 - 30°C/min (increment: 0.1°C/min), ±0.1% temperature control precision
Temperature Control Accuracy ≤0.1ºC
Dimensions (L×W×H) 500mm × 430mm × 440mm
Net Weight 50 kg